DSP-100/2000

Users Manual

Pressing @ View Plot produces the RL plot screen. Figure 3-6 shows an example of the screen and Table 3-9 describes the items on the screen.

6

5

4

1

2

3

gc13c.eps

Item

1

2

3

4

5

6

Figure 3-6. The RL Plot Screen

Table 3-9. Items on the RL Plot Screen

Description

The cable pair relevant to the plot.

Frequency range in MHz of the RL test.

Margin is the difference between the limit and measured values plotted at the cursor’s position. Use LRto move the cursor left or right. If you move the cursor beyond the highest test frequency specified by the selected test standard, the readout shows the RL value at the cursor’s position.

The limits for RL, as defined by the selected test standard.

Decibels of RL for the cable pair.

The measured RL for the cable pair.

RL@REMOTE (Model DSP-2000)

The RL@REMOTE test is identical to the RL test, except that the RL values are measured from the remote end of the cable.

3-18

Page 66
Image 66
Fluke DSP-100 user manual RL@REMOTE Model DSP-2000