Triggering the Oscilloscope with the Timing Analyzer

Add Oscilloscope Waveform to the Timing Waveforms

Add Oscilloscope Waveform to the Timing Waveforms

To see the correlation between the timing analyzer and the oscilloscope, you must insert the scope display into the timing waveform display.

1Press the Wform MENU key to display the Waveform Machine1 menu.

2Highlight the large field at the left side of the display. The TCOUNT label should be displayed in the large field.

If the TCOUNT label is not in the large field, then you need to add it. To add the TCOUNT label, highlight the large field, then press the Select key twice. Choose Insert, Analyzer, TCOUNT, then Sequential in the pop-up menus.

3Using the knob, place the cursor on the TCOUNT 7 label.

4Press the Select key.

5Choose Insert, Scope, then C1 (oscilloscope channel 1) in the pop-up menus.

The oscilloscope waveform is added to the analyzer display.

 

Adding Oscilloscope Waveform to the Timing Waveforms

Result

Both the timing waveform for TCOUNT and the oscilloscope waveform will be

 

displayed when you run the analyzer.

8-13