Introduction to State Analysis

Set Up Level 2 of the State Trigger Specification

Set Up Level 2 of the State Trigger Specification

In sequence level 1 the analyzer stores data before the trigger point. For sequence level 2, you will tell the analyzer to store any data once the trigger pattern is found. This will fill the rest of the acquisition memory.

1Highlight the 2 field in the State Sequence Levels box. Press the Select key.

2Highlight the field to the right of Store, and press the Select key. Highlight anystate in the pop-up menu, and press the Done key.

3Press the Done key to close the Sequence Level 2 pop-up menu.

Once the trigger pattern is found, the logic analyzer will store any data that it samples.

 

Setting Up Level 2 of the State Trigger Specification

Result

The State Trigger Specification is now:

 

Level 1

- While storing "anystate", TRIGGER on "b" 1 time

 

Level 2

- Store "anystate"

If your trigger specification does not match this one, repeat the exercises, starting with "Modify Channel Assignments" on page 3–7.

3-10