Comparing State Traces

Check the Configuration

Check the Configuration

Either you have just completed chapter 3, "Introduction to State Analysis", or you have loaded the configuration file CH_04.

1If you are using an HP 1660 series analyzer go to step 2. If you are using an HP 1670 series analyzer, put the analyzer into State Compare mode before checking the configurations.

To do this, press the Config MENU key. In the Analyzer 1 box, highlight the field to the right of Type, press the Select key, and then choose State Compare from the pop-up menu.

2Check the configuration.

In the Config menu, Analyzer 1 is set to State mode for the HP 1660s and State Compare for the HP 1670s.

In the Format menu, the Master Clock field is displaying J ↓.

In the Format menu, the SCOUNT label is turned on with the lower eight channels of pod A1 assigned to it.

In the Trigger menu SCOUNT’s term "a" is assigned the value FF (hex).

4-4