Using the Oscilloscope

Measure the Clock Period Manually

Measure the Clock Period Manually

Time and voltage measurements can be made manually by turning the

T Markers field on.

1Press the Marker MENU key. Marker is printed in white on the MENU key.

2Highlight the T Markers Off field and press the Select key. Highlight On in the pop-up menu, and press the Select key.

3Highlight the Trig to X field. Using the knob, move the X Marker to the lowest point on the negative pulse.

You may need to expand or compress the waveform, using the s/Div field to see a complete period of the waveform on the display.

4Highlight the Trig to O field. Using the knob move the O Marker to the same point on the next negative edge after the X Marker.

5Read the period of the clock in the Tx to To field.

Read the clock period in this field.

Highlight this field and select On.

Highlight this field and position the X Marker.

Highlight this field and position the O Marker.

Measuring the Clock Period Manually with the X and O Markers

7-8