Triggering the Oscilloscope with the Timing Analyzer

Align the Glitch Displayed by the Oscilloscope and Analyzer

4Press the Trigger MENU key.

5Highlight the Arming Control field and press the Select key.

6With the Skew field highlighted, press the Select key.

7With the Analyzer field highlighted, press the Select key.

8Use the knob to change the units to ns, and type in the time delay you measured with the Trig to O marker.

9Press the Done key until you return to the Trigger Menu.

10Highlight the Group Run field, and press the Select key. Highlight Single in the pop-up menu, and press the Select key.

The timing waveform is now lined up with the oscilloscope waveform.

 

Aligning the Glitch Displayed by the Oscilloscope and Analyzer

Result

The analyzer display is time-aligned with the scope display.