Advanced State Triggering
Set Up Level 2 of the State Trigger Specification
Set Up Level 2 of the State Trigger Specification
The second term you will look for is the range of values 050 through 058, which you defined as "In range1". While you are storing these values, you also want to find and store term "e" with the value 059.
1Highlight the 2 field in the State Sequence Levels box of the Trigger menu. Press the Select key.
2Highlight the field to the right of While storing, and press the Select key. Highlight In_Range1 in the
3Highlight the field to the right of Then find, and press the Select key. Highlight e in the
4Press the Done key to close the Sequence Level 2
In_Range1 tells the analyzer to store the values 050 through 058.
While storing In_Range1, the analyzer looks for and stores "e" one time.
| Setting Up Level 2 of the State Trigger Specification |
Result | Sequence Level 2 is set to store the "In_Range1" values 050 through 058 and |
| then find and store the term "e" value 059. |