Triggering the Oscilloscope with the Timing Analyzer

Align the Glitch Displayed by the Oscilloscope and Analyzer

Align the Glitch Displayed by the Oscilloscope and

Analyzer

You can align the glitch displayed by the analyzer and scope by measuring the distance between the falling edge on the timing display with the falling edge of the scope display. Then you use the skew feature to move the timing waveform over so that the two displays are aligned.

1Highlight the Markers Off field, and press the Select key. Highlight Time in the pop-up menu, and press the Select key.

2Highlight the Trig to O field, and use the knob to place the O marker on the falling edge of the glitch shown by the scope.

3Take note of the time delay in ns between the trigger point and the O marker. This is the time delay you will use to realign the two displays.

The time delay.

Measuring the Time Delay