Triggering the Oscilloscope with the Timing Analyzer
Set Up the Timing Trigger Specification
Set Up the Timing Trigger Specification
You will now tell the analyzer to trigger on the glitch the first time it occurs.
1Highlight the 1 field in the Timing Sequence Levels box, and press the
Select key.
2Highlight the ">" field, and press the Select key. Highlight Occurs in the
3Highlight the field next to Trigger on, and press the Select key. Highlight Edge 1 in the
4Highlight the field next to Else on, and press the Select key. Highlight no state in the
5Press the Done key to close the Sequence Level 1
The trigger term "Edge1" is set for the analyzer to look for the glitch on bit 7.
| Setting Up the Trigger Specification |
Result | Your trigger specification now shows: Level 1 - TRIGGER on "Edge1" 1 time. |