Triggering the Oscilloscope with the Timing Analyzer

Set Up the Timing Trigger Specification

Set Up the Timing Trigger Specification

You will now tell the analyzer to trigger on the glitch the first time it occurs.

1Highlight the 1 field in the Timing Sequence Levels box, and press the

Select key.

2Highlight the ">" field, and press the Select key. Highlight Occurs in the pop-up menu, and press the Select key.

3Highlight the field next to Trigger on, and press the Select key. Highlight Edge 1 in the pop-up menu, and press the Select key.

4Highlight the field next to Else on, and press the Select key. Highlight no state in the pop-up menu, and press the Select key.

5Press the Done key to close the Sequence Level 1 pop-up menu.

The trigger term "Edge1" is set for the analyzer to look for the glitch on bit 7.

 

Setting Up the Trigger Specification

Result

Your trigger specification now shows: Level 1 - TRIGGER on "Edge1" 1 time.

8-11