Advanced State Triggering
Check the Trigger Specification
Check the Trigger Specification
Now you can check the trigger specification by scrolling through the five sequence levels. You did not have to set up Level 5 because you want the analyzer to trigger and store "anystate". The last level of a trigger specification has this as the default.
1Highlight the "State Sequence Levels" field centered above the sequence specification.
2Turn the knob clockwise to scroll down through the sequence levels.
Your trigger specification should look like the one below:
•Level 1 - While storing "no state"; Find "a" 1 time
•Level 2 - While storing "In_Range1"; Then find "e" 1 time
•Level 3 - While storing "no state"; Then find "b" 1 time
•Level 4 - While storing "c + In_Range1 + d" ; TRIGGER on "e" 1 time
•Level 5 - Store "anystate"
If your trigger specification is not correct, repeat the appropriate exercises on the previous pages to correct the sequence levels that do not match.