Mixed Mode - Correlation of State and Timing Data

Clear the Timing Analyzer Trigger

Clear the Timing Analyzer Trigger

Clearing the timing analyzer trigger sets the specification to "don’t care". By doing this, the timing analyzer will trigger as soon as it is armed by the state analyzer.

1Press the Trigger MENU key.

2Highlight the Modify Trigger field on the right side of the screen, and press the Select key.

3Highlight the Clear Trigger field, press the Select key, and choose All from the pop-up menu.

Result

The timing analyzer trigger specification is set to a "don’t care" state.

Set Up the State Trigger Specification

1Press the Trigger MENU key. Highlight Trigger MACHINE 1 in the pop-up menu, and press the Select key. The Trigger MACHINE 1 menu is now displayed. If this is not the case, press the Trigger MENU key again, then choose Trigger MACHINE 1 from the pop-up menu.

2Highlight the 1 field in the State Sequence Levels box of the Trigger menu. Press the Select key.

3Highlight the field to the right of While storing, and press the Select key. Highlight anystate in the pop-up menu, and press the Select key.

4Press the Done key to close the Sequence Level 1 pop-up menu.

Result

The state analyzer trigger specification is now:

 

Level 1

- While storing "anystate", TRIGGER on "a" 1 time

 

Level 2

- Store "anystate"

The trigger term "a" is assigned the hex value FF.

5-7