Mixed Mode - Correlation of State and Timing Data
Clear the Timing Analyzer Trigger
Clear the Timing Analyzer Trigger
Clearing the timing analyzer trigger sets the specification to "don’t care". By doing this, the timing analyzer will trigger as soon as it is armed by the state analyzer.
1Press the Trigger MENU key.
2Highlight the Modify Trigger field on the right side of the screen, and press the Select key.
3Highlight the Clear Trigger field, press the Select key, and choose All from the
Result | The timing analyzer trigger specification is set to a "don’t care" state. |
Set Up the State Trigger Specification
1Press the Trigger MENU key. Highlight Trigger MACHINE 1 in the
2Highlight the 1 field in the State Sequence Levels box of the Trigger menu. Press the Select key.
3Highlight the field to the right of While storing, and press the Select key. Highlight anystate in the
4Press the Done key to close the Sequence Level 1
Result | The state analyzer trigger specification is now: | |
| Level 1 | - While storing "anystate", TRIGGER on "a" 1 time |
| Level 2 | - Store "anystate" |
The trigger term "a" is assigned the hex value FF.