Triggering the Oscilloscope with the Timing Analyzer
Change the Oscilloscope Trigger
Change the Oscilloscope Trigger
You will be using the timing analyzer to find the glitch and then the oscilloscope to display it. This correlation of timing and scope is done by arming the scope with the timing analyzer. To capture the glitch on the scope, you want it to capture data immediately after the glitch is captured by the timing analyzer.
1Press the Trigger MENU key.
2Highlight the Mode/Arm Edge field at the left side of the display, and press the Select key.
3Highlight Immediate in the
Setting this field to Immediate tells the oscilloscope to trigger immediately after the timing analyzer triggers.
Changing the Oscilloscope Trigger
Result | The oscilloscope is set to capture data after the timing analyzer triggers on |
| the glitch. |