General Characteristics

Power Requirements

Settling Time:

300 μS (+full scale to –full scale step, single channel, to accuracy listed previously)

Isolation (for isolated channels):

42 Vdc/42 V peak (channel-to-chassis or channel-to-channel)

Synchronization:

Software commands, External TTL trigger, or VXIbus TTL Trigger lines Single channel or multiple channel

VXI Specification:

Device Type: A16 or A24, D16, slave only, register-based

Current Required (Amps)

Configuration

+ 5 V

+ 5 V

+ 12 V

+ 12 V

+ 24 V

+ 24 V

– 24 V

– 24 V

 

DC

Dynamic

DC

Dynamic

DC

Dynamic

DC

Dynamic

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8 Channel

0.70

0.01

0.04

0.01

0.44

0.01

0.44

0.01

Non-Isolated

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8 Channel

0.75

0.01

0.04

0.01

0.88

0.01

0.00

0.00

Isolated

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16 Channel

1.25

0.01

0.04

0.01

0.88

0.01

0.88

0.01

Non-Isolated

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16 Channel

1.40

0.01

0.04

0.01

1.60

0.01

0.00

0.00

Isolated

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cooling For a 10°C rise

Requirements

 

 

Δ P

Avg

 

Air Flow

Watts/slot

 

 

 

 

 

 

 

 

8 Channel

2.0

0.10 mm H2O

25.4

configurations

liters/second

 

 

 

 

 

 

16 Channel

3.9

0.18 mm H2O

49.0

configurations

liters/second

 

 

 

 

 

 

 

 

 

 