id
Error Messages
|
| Table |
|
|
|
Code | Error Messages | Potential Cause(s) |
|
|
|
|
|
|
Invalid character data | An invalid SCPI keyword was used. | |
|
|
|
Trigger ignored | A trigger was received and no channels were in the wait for | |
|
| trigger state. |
|
|
|
Init ignored | An INITiate was received and one or more channels were | |
|
| already in the wait for trigger state. |
|
|
|
Setting Conflict | A command was received for a mode not set. For example, a | |
|
| SOURn:VOLT command was received and the channel is set to |
|
| CURR mode. |
|
|
|
Data out of range | <n> in SCPI keywor s not in the range of 1 to 16. | |
|
| <value> is out of range. |
|
| <mask> is out of range. |
|
| <number> is out of range. |
|
|
|
Illegal Parameter Value | Value received is not in the legal range for the parameter. | |
|
|
|
Hardware Error | A hardware failure has been detected. The module may require | |
|
| replacement or repair. |
|
|
|
+2001 | Invalid Channel Number | <n> in SCPI keyword is not valid for the module configuration. |
|
| For example <n> is set to 11 on an |
|
|
|
+3000 | Trigger too fast | The trigger was received before the module completed the last |
|
| command. |
|
|
|
+3002 | Self Test Failed | A hardware failure was detected during power up |
|
| Additional |
|
| beginning on page 166. |
|
|
|
+3004 | Multiple attempts to erase | Attempts to erase |
| command failed. May indicate a hardware failure or the CAL | |
|
| Store Enable Jumper is not in the correct position. |
|
|
|
+3005 | Multiple attempts to program | Attempts to write to |
| command failed. May indicate a hardware failure or the CAL | |
|
| Store Enable Jumper is not in the correct position. |
|
|
|
+3006 | CAL Store Enable Jumper not set | A CAL:STOR command was received and the CAL Store |
| properly | Enable Jumper is not set to the CAL position. |
|
|
|
+3007 | Checksum error on | Calibration constants in |
| Memory | Recalibration is needed. |
|
|
|
|
|
|
151 HP E1418A Error Messages | Appendix C |