id

Error Messages

 

 

Table C-2. Error Messages

 

 

 

Code

Error Messages

Potential Cause(s)

 

 

 

 

 

 

–141

Invalid character data

An invalid SCPI keyword was used.

 

 

 

–211

Trigger ignored

A trigger was received and no channels were in the wait for

 

 

trigger state.

 

 

 

–213

Init ignored

An INITiate was received and one or more channels were

 

 

already in the wait for trigger state.

 

 

 

–221

Setting Conflict

A command was received for a mode not set. For example, a

 

 

SOURn:VOLT command was received and the channel is set to

 

 

CURR mode.

 

 

 

–222

Data out of range

<n> in SCPI keywor s not in the range of 1 to 16.

 

 

<value> is out of range.

 

 

<mask> is out of range.

 

 

<number> is out of range.

 

 

 

–224

Illegal Parameter Value

Value received is not in the legal range for the parameter.

 

 

 

–240

Hardware Error

A hardware failure has been detected. The module may require

 

 

replacement or repair.

 

 

 

+2001

Invalid Channel Number

<n> in SCPI keyword is not valid for the module configuration.

 

 

For example <n> is set to 11 on an 8-channel configuration.

 

 

 

+3000

Trigger too fast

The trigger was received before the module completed the last

 

 

command.

 

 

 

+3002

Self Test Failed

A hardware failure was detected during power up self-test.

 

 

Additional self-test information is given in Appendix E,

 

 

beginning on page 166.

 

 

 

+3004

Multiple attempts to erase

Attempts to erase non-volatile memory using the CAL:STOR

 

non-volatile Memory failed

command failed. May indicate a hardware failure or the CAL

 

 

Store Enable Jumper is not in the correct position.

 

 

 

+3005

Multiple attempts to program

Attempts to write to non-volatile memory using the CAL:STOR

 

non-volatile Memory failed

command failed. May indicate a hardware failure or the CAL

 

 

Store Enable Jumper is not in the correct position.

 

 

 

+3006

CAL Store Enable Jumper not set

A CAL:STOR command was received and the CAL Store

 

properly

Enable Jumper is not set to the CAL position.

 

 

 

+3007

Checksum error on non-volatile

Calibration constants in non-volatile memory are corrupted.

 

Memory

Recalibration is needed.

 

 

 

 

 

 

151 HP E1418A Error Messages

Appendix C