HP E1418A manual Store the NEW Calibration Constants

Models: E1418A

1 187
Download 187 pages 37.67 Kb
Page 164
Image 164

err=viScanf(dac, “%d”, &result);

/* Enter integer returned */

 

/*, returns a 1 when ready to calibrate */

if(err < VI_SUCCESS) err_handler(dac, err);

}

 

result = 1;

 

while (result !=0 && result !=2)

 

{

 

meas_volt=get_voltage ();

/* Get the voltage measurement */

err=viPrintf(dac, “CAL%d:VAL:CURR? %f\n”,i,meas_volt); if(err < VI_SUCCESS) err_handler(dac, err);

err=viScanf(dac, “%d”, &result);

/* Enter integer returned */

 

/* returns a 0 when calibration complete */

if(err < VI_SUCCESS) err_handler(dac, err);

if (result == 5)

/* Calibration aborted */

{

 

printf (“Calibration for channel %d aborted\n\n”,i);

result = 0;

 

}

 

}

 

}

 

/*** STORE THE NEW CALIBRATION CONSTANTS

***/

err=viPrintf(dac, “CAL:STOR;*OPC?\n”);

/* Store the new cal constants */

if(err < VI_SUCCESS) err_handler(dac, err);

 

err=(viScanf(dac,"%d",&result);

 

if(err < VI_SUCCESS) err_handler(dac, err);

sys_err(dac);

 

 

 

}

 

Appendix D

Voltage/Current Output Adjustment 164