DIAGnostic
DIAGnostic:OUTPut:ALL[:STATe]
DIAGnostic:OUTPut:ALL[:STATe] <state> opens or closes all channel output relays at once.
Parameters
Parameter | Parameter | Range of | Default |
Name | Type | Values |
|
|
|
|
|
<state> | boolean | OFF ON 0 1 | 0 |
|
|
|
|
Comments ∙ This command performs the same function as the OUTPutn[:STATe] command, but enables or disables all outputs simultaneously.
∙Related Commands: OUTPutn[:STATe]
DIAGnostic:SOURcen:FUNCtion:MODE
DIAGnostic:SOURcen:FUNCtion:MODE <mode> changes the output mode of the
channel without disabling the channel output.
Parameters
| Parameter | Parameter | Range of | Default |
| Name | Type | Values |
|
|
|
|
|
|
| SOURcen | keyword | 1 to 8 or 1 to 16 | 1 |
|
|
|
|
|
| <mode> | discrete | CURRent or VOLTage | none |
|
|
|
|
|
Comments ∙ In the keyword, SOURcen, the valid range of n is: |
| |||
| 1 to 8 for | E1418A, E1418A Option 002 |
| |
| 1 to 16 for | E1418A Option 001, E1418A Option 003. |
|
Values of n outside these ranges will generate an error.
∙Use this command with caution. The output voltage or current is not predictable during the mode change and transients will occur.
∙This command performs the same function as the SOURcen:FUNCtion:MODE command, but does not disable the output during the mode change. This command can reduce the settling time required to disable and enable the outputs.
∙Following a mode change with this command, the appropriate calibration constants are used, the channel mode is changed, and the output is set to either 0 V or 0 A.
∙This command will generate an error if the trigger system is waiting for a trigger event when this command is received.
∙This command is affected by the state of the P/J jumper on the terminal module. If the P/J jumper is not set to Program Selectable, this command will generate an error. This jumper is described on page 31.
∙Related Commands: SOURcen:FUNCtion:MODE
78 HP E1418A SCPI Command Reference | Chapter 3 |