CALibration
CALibrationn:CONFigure:VOLTage?
CALibrationn:CONFigure:VOLTage? configures the channel for voltage mode calibration and returns a 0 or 1 to indicate the channel’s readiness for calibration.
Parameters
Parameter | Parameter | Range of | Default |
Name | Type | Values |
|
|
|
|
|
CALibrationn | keyword | 1 to 8 or 1 to 16 | 1 |
|
|
|
|
Returns: an unsigned
Comments ∙ In the keyword, CALibrationn, the valid range of n is:
1 to 8 for | E1418A, E1418A Option 002 |
1 to 16 for | E1418A Option 001, E1418A Option 003. |
Values of n outside these ranges will generate an error.
∙This command is used together with the CALibrationn:VALue:VOLTage? query to perform voltage mode calibration.
∙The calibration is performed either at the CAL output terminals or at an individual channel. The choice is made using the DIAGnostic:CALibration:OUTPut[:MODE] command.
–When the DIAG:CAL:OUTP:MODE CALBus command (also the reset state)
is used, the output relay routes the calibration voltage to the CAL output terminals (and opens all channel output terminals).
–When the command DIAG:CAL:OUTP:MODE CHANnel is used, the output relay routes the calibration voltage to the channel terminals (the CAL terminals are not connected).
The DIAGnostic:CALibration:OUTPut[:MODE] command must be sent to the module before the CALibrationn:CONFigure:VOLTage? query.
∙This query will generate an error if the trigger system is waiting for a trigger event when this command is received.
∙This query returns a 0 to indicate that the channel is not ready for calibration, and a 1 to indicate the channel is ready for calibration.
∙This command is affected by the state of the program jumper for this channel on the terminal module. To use this command:
–the P/J jumper must be set to program selectable or, if set to jumper selectable,
–the V/I jumper must be set to the voltage mode. The jumpers are described on page 31.
∙Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154, for additional information about this command.
∙Related Commands: CALibrationn:VALue:VOLTage?, DIAGnostic:CALibration:OUTPut[:MODE]
72 HP E1418A SCPI Command Reference | Chapter 3 |