CALibration

A returned value of 5 (calibration aborted) is given if either:

a.The command is received when the module is in voltage calibration or resistance calibration mode, or

b.The calibration in process has entered more than 50 values without a successful calibration.

If the internal calibration resistor is not calibrated, the channel is not marked as calibrated following the completion of calibration (see the CALibration:CONDition? query). The new calibration constants are used, however.

When the channel current calibration is completed, the new calibration constants are used by the module, but not stored in non-volatile memory.

If a calibration is interrupted before completion (that is, by another command being received), no constants will be changed.

Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154, for additional information about this command.

Related Commands: CALibration:STORe, CALibrationn:CONFigure:CURRent?

CALibrationn:VALue:VOLTage?

CALibrationn:VALue:VOLTage? <value> enters a measured calibration value and returns a value indicating the present state of the voltage calibration process.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

 

 

 

 

 

CALibrationn

keyword

1 to 8 or 1 to 16

1

 

 

 

 

<value>

numeric (float 32)

0 to 19.2 V

none

 

 

 

 

Returns: unsigned 16-bit integer (uint16).

Comments In the keyword, CALibrationn, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

This command is used together with the CALibrationn:CONFigure:VOLTage? query to perform voltage mode calibration. Following the CALn:CONF:VOLT? query, this command is used repetitively until the return value indicates that the calibration is complete.

The <value> parameter is the measured output voltage. A minimum of 9 measurements must be taken to calibrate a voltage output channel (a maximum of 50 measurements may be required).

74 HP E1418A SCPI Command Reference

Chapter 3