CALibration

CALibration:STORe

CALibration:STORe loads the stored non-volatile memory calibration constants with the calibration constants currently in use by the module.

Parameters None.

Comments This command sets the calibration constants in non-volatile memory if the CAL Store Enable Jumper is properly set (see page 156). Non-volatile memory has finite life, based upon the number of writes. Use this command sparingly.

At power-on, calibration constants stored in non-volatile memory are loaded into the module where they are used during output operations.

When this command is received, the checksum is verified using the constants currently in use by the module. If the checksum does not verify, the non-volatile memory is not updated and an error is generated.

Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154, for additional information about this command.

Related Commands: CALibration:RESet

CALibration:TEST?

CALibration:TEST? performs an internal rough check of the calibration of each channel and returns an integer indicating the channel calibration condition. Calibration is checked to approximately ± 2%.

Parameters Returns: unsigned 16-bit integer (uint16).

Comments Following the completion of this query all outputs are disabled and set to 0 Volts or 0 Amps based on the channel mode.

This query returns a 16-bit integer indicating the calibration status of each channel.

Bits set to 0 indicate no calibration errors were found.

Bits set to 1 indicate a channel with questionable calibration.

Bits 0–15 corre spond to channels 1–16, respectively.

If the trigger system is waiting for a trigger when this command is received, an error is generated.

Chapter 3

HP E1418A SCPI Command Reference 69