APPLy

APPLy

 

The APPLy subsystem enables and outputs a current or voltage on the specified channel.

Subsystem Syntax

APPLyn

 

:CURRent <value MIN MAX DEF>

 

:VOLTage <value MIN MAX DEF>

APPLyn:CURRent

APPLyn:CURRent <value MIN MAX DEF> outputs a current on the specified channel, n.

Parameters

 

Parameter

Parameter

Range of

Default

 

Name

Type

Values

 

 

 

 

 

 

 

APPLyn

keyword

1 to 8 or 1 to 16

1

 

 

 

 

 

 

<value>

numeric (float 32) or

-0.02 Amps to +0.02 Amps or

0

 

 

discrete

MIN MAX DEF

 

 

 

 

 

 

Comments In the keyword, APPLyn, the valid range of n is:

 

 

1 to 8 for

E1418A, E1418A Option 002

 

 

1 to 16 for

E1418A Option 001, E1418A Option 003.

 

Values of n outside these ranges will generate an error.

The <value> parameter can range from -0.02 to +0.02 Amps. MIN sets -0.02 Amps output.

MAX sets +0.02 Amps output. DEF sets 0 Amps output.

This command is affected by the state of the P/J jumper for this channel on the terminal module. To use the APPLyn:CURRent command:

the P/J jumper must be set to program selectable or, if set to jumper selectable,

the V/I jumper must be set to the current mode. These jumpers are described on page 31.

Equivalent Commands: SOURcen:FUNCtion:MODE CURR,

SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude], OUTPutn[:STATe] ON

*RST Condition:

If the P/J jumper is set to program selectable:

output mode is set to voltage, 0 volts output, output relay open.

If the P/J jumper is set to jumper selectable:

output mode is set by the V/I jumper, 0 volts or amps, output relay open.

64 HP E1418A SCPI Command Reference

Chapter 3

Page 64
Image 64
HP E1418A manual APPLynCURRent