APPLy
APPLyn:VOLTage
APPLyn:VOLTage <value MIN MAX DEF> outputs a voltage on the specified channel, n.
Parameters
| Parameter | Parameter | Range of | Default |
| Name | Type | Values |
|
|
|
|
|
|
| APPLyn | keyword | 1 to 8 or 1 to 16 | 1 |
|
|
|
|
|
| <value> | numeric (float 32) or | 0 | |
|
| discrete |
|
|
|
|
|
|
|
Comments ∙ In the keyword, APPLyn, the valid range of n is: |
| |||
| 1 to 8 for | E1418A, E1418A Option 002 |
| |
| 1 to 16 for | E1418A Option 001, E1418A Option 003. |
|
Values of n outside these ranges will generate an error.
∙The <value> parameter can range from
MAX sets +16 Volts output. DEF sets 0 Volts output.
∙This command is affected by the state of the program jumper on the terminal module for this channel. To use the APPLyn:VOLT command:
–the P/J jumper must be set to program selectable or, if set to jumper selectable,
–the V/I jumper must be set to the voltage mode. This jumper is described on page 31.
∙Equivalent Commands: SOURcen:FUNCtion:MODE VOLT,
SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude], OUTPutn[:STATe] ON
∙*RST Condition:
–If the P/J jumper is set to program selectable:
output mode is set to voltage, 0 volts output, output relay open.
–If the P/J jumper is set to jumper selectable:
output mode is set by the V/I jumper, 0 volts or amps, output relay open.
Chapter 3 | HP E1418A SCPI Command Reference 65 |