SOURce

SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude]?

SOURcen:CURRent[:LEVel]:TRIGgered][:AMPLitude]? [<MIN MAX DEF>

queries the last value written by either

SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude] or

SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude].

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

 

 

 

 

 

SOURcen

keyword

1 to 8 or 1 to 16

1

 

 

 

 

<MIN MAX DEF>

discrete (optional)

MIN MAX DEF

none

 

 

 

 

Returns: 32-bit floating point number (float 32).

Comments In the keyword, SOURcen, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

If no parameter is given, this query returns the last value set on channel n by the last SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude] or SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude] command. The value returned is the value last programmed on that channel, not an actual measurement of the hardware output value.

If a parameter is given,

MIN returns -0.02 Amps.

MAX returns +0.02 Amps.

DEF returns 0 Amps.

This command generates an error if channel n is in the voltage output mode.

Related Commands: SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude], SOURcen:FUNCtion:MODE CURRent

88 HP E1418A SCPI Command Reference

Chapter 3

Page 88
Image 88
HP E1418A manual SOURce SOURcenCURRentLEVelTRIGgeredAMPLitude?, SOURcenCURRentLEVelTRIGgeredAMPLitude? MIN MAX DEF