SOURce
SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude]?
SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude]? [<MIN MAX DEF>
queries the last output value written by
SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude].
Parameters
Parameter | Parameter  | Range of  | Default  | 
Name  | Type  | Values  | 
  | 
  | 
  | 
  | 
  | 
SOURcen  | keyword  | 1 to 8 or 1 to 16  | 1  | 
  | 
  | 
  | 
  | 
<MIN MAX DEF>  | discrete (optional)  | MIN MAX DEF  | none  | 
  | 
  | 
  | 
  | 
Returns: 
Comments ∙ In the keyword, SOURcen, the valid range of n is:
1 to 8 for  | E1418A, E1418A Option 002  | 
1 to 16 for  | E1418A Option 001, E1418A Option 003.  | 
Values of n outside these ranges will generate an error.
∙If no parameter is given, this query returns the source current set on channel n by the last SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude] command. The value returned is the value last programmed on that channel, not an actual measurement of the hardware output value.
∙If a parameter is given,
MIN returns 
MAX returns +0.02 Amps.
DEF returns 0 Amps.
∙This command generates an error if channel n is in the voltage output mode.
∙Related Commands: SOURcen:CURRent[:LEVel][:IMMediate][:AMPLitude], SOURcen:FUNCtion:MODE CURRent
86 HP E1418A SCPI Command Reference | Chapter 3  |