Combining

You can combine channels in series to obtain higher output voltages or in

Channels

parallel to obtain higher output currents.

Channels in Series

All channels connected in series MUST be configured as isolated

 

channels (with isolated plug-on modules installed).

 

Up to 3 channels may be combined, but the output voltage MUST

 

NOT exceed 42 Vdc/42 Vpeak.

Channels in Parallel

Channels may be either Isolated or Non-Isolated.

 

Up to 16 channels may be combined.

 

Figure 1-15 shows how to combine voltage channels (in series) and current

 

channels (in parallel).

CAUTIONS

 

All SERIES CONNECTED output channels MUST be configured as

 

ISOLATED OUTPUTS.

 

When combining output voltage channels, be sure not to exceed the

 

maximum module isolation rating of 42 Vdc or 42 Vpeak.

 

 

Figure 1-15. Combined Channels Output Connections