STATus
STATus:OPERation:ENABle?
STATus:OPERation:ENABle? returns the value of the mask set for the Operation
Status Register.
Parameters Returns:
STATus:OPERation[:EVENt]?
STATus:OPERation[:EVENt]? returns the value of the contents of the Operation
Status Event Register. Reading the register clears all bits in the register.
Parameters Returns:
Comments ∙ Bit 0 of this register (decimal 1) is the only defined bit in this register.
∙This bit is set if the CALibration:CONFigure:RESistance?,
CALibrationn:CONFigure:VOLTage?, or CALibrationn:CONFigure:CURRent? command has been received and the calibration process is not yet complete.
∙*RST Condition: no effect.
∙*CLS clears the contents of the Operation Status Register.
∙STATus:PRESet does not affect the contents of the Operation Status Register, but does disable reporting the Summary bit to the Status Byte Register.
STATus:PRESet
STATus:PRESet presets all Enable Registers to 0 (zero). The command has no effect on Condition or Event Registers.