CALibration

CALibrationn:VALue:CURRent?

CALibrationn:VALue:CURRent? <value> enters a measured calibration value and returns a value indicating the present state of the current calibration process.

Parameters

Parameter

Parameter

Range of

Default

Name

Type

Values

 

 

 

 

 

CALibrationn

keyword

1 to 8 or 1 to 16

1

 

 

 

 

<value>

numeric (float 32)

0.0 to 6.0 V

none

 

 

 

 

Returns: unsigned 16-bit integer (uint16).

Comments In the keyword, CALibrationn, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

This command is used together with the CALibrationn:CONFigure:CURRent? query to perform voltage mode calibration. Following the CALn:CONF:CURR? query, this command is used repetitively until the return value indicates that the calibration is complete.

The <value> parameter is the measured output voltage. A minimum of

9 measurements must be taken to calibrate a channel for current mode output (a maximum of 50 measurements may be required).

This query returns one of the following values:

Return

Meaning

Value

0

Calibration is complete.

 

New constants are in RAM.

 

Resistance is calibrated.

1

More measurements are needed.

 

Resistance is calibrated.

 

 

2

Calibration is complete.

 

New constants are in RAM.

 

Resistance is not calibrated.

3

More Measurements are needed.

 

Resistance is not calibrated.

 

 

5

Calibration aborted.

Chapter 3

HP E1418A SCPI Command Reference 73