SOURce

SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude]

SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude] <value MIN MAX DEF>

outputs the current set in value at channel n when the trigger conditions are satisfied and the output relay is closed. Use this command to synchronize multiple channel outputs.

Parameters

 

Parameter

Parameter

Range of

Default

 

Name

Type

Values

 

 

 

 

 

 

 

SOURcen

keyword

1 to 8 or 1 to 16

1

 

 

 

 

 

 

<value>

numeric (float 32) or

-0.02 Amps to +0.02 Amps or

0

 

 

discrete

MIN MAX DEF

 

 

 

 

 

 

Comments In the keyword, SOURcen, the valid range of n is:

 

 

1 to 8 for

E1418A, E1418A Option 002

 

 

1 to 16 for

E1418A Option 001, E1418A Option 003.

 

Values of n outside these ranges will generate an error.

The <value> parameter is a floating point 32-bit number in the range of -0.02 A to +0.02 A.

MIN sets -0.02 Amps. MAX sets +0.02 Amps. DEF sets 0 Amps.

This command sets channel n to output the current specified following a valid trigger condition. Following this command, the trigger system must be enabled using an INITiate command. See the TRIGger subsystem discussion on page 54 for additional information.

If the module is in a wait-for-trigger state (INIT has been received) when this command is received, an error is generated.

The output state of the channel must be enabled and the trigger event satisfied for the current to appear at the output terminals.

If the channel is in the voltage mode when this command is received, an error is generated.

Related Commands: SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude]?, SOURcen:FUNCtion:MODE CURRent, OUTPutn[:STATe]

Chapter 3

HP E1418A SCPI Command Reference 87