SOURce
SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude]
SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude] <value MIN MAX DEF>
outputs the current set in value at channel n when the trigger conditions are satisfied and the output relay is closed. Use this command to synchronize multiple channel outputs.
Parameters
| Parameter | Parameter | Range of | Default |
| Name | Type | Values |
|
|
|
|
|
|
| SOURcen | keyword | 1 to 8 or 1 to 16 | 1 |
|
|
|
|
|
| <value> | numeric (float 32) or | 0 | |
|
| discrete | MIN MAX DEF |
|
|
|
|
|
|
Comments ∙ In the keyword, SOURcen, the valid range of n is: |
| |||
| 1 to 8 for | E1418A, E1418A Option 002 |
| |
| 1 to 16 for | E1418A Option 001, E1418A Option 003. |
|
Values of n outside these ranges will generate an error.
∙The <value> parameter is a floating point
MIN sets
∙This command sets channel n to output the current specified following a valid trigger condition. Following this command, the trigger system must be enabled using an INITiate command. See the TRIGger subsystem discussion on page 54 for additional information.
∙If the module is in a
∙The output state of the channel must be enabled and the trigger event satisfied for the current to appear at the output terminals.
∙If the channel is in the voltage mode when this command is received, an error is generated.
∙Related Commands: SOURcen:CURRent[:LEVel]:TRIGgered[:AMPLitude]?, SOURcen:FUNCtion:MODE CURRent, OUTPutn[:STATe]
Chapter 3 | HP E1418A SCPI Command Reference 87 |