CALibration

This query returns one of the following values:

Return

ValueMeaning

0Calibration is complete.

New constants are in RAM.

1More measurements are needed.

5Calibration aborted.

A return value of 5 (calibration aborted) is given if either:

a.The command is received when the module is in current calibration or resistance calibration mode, or

b.The calibration in process has entered more than 50 values without a successful calibration.

When the channel voltage calibration is completed, the new calibration constants are used by the module, but not stored in non-volatile memory.

If a voltage calibration is interrupted before completion (that is, by another command being received), no constants are changed.

Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154, for additional information about this command.

Related Commands: CALibration:STORe, CALibrationn:CONFigure:VOLTage?

Chapter 3

HP E1418A SCPI Command Reference 75

Page 75