CALibration

CALibration:CONFigure:RESistance?

CALibration:CONFigure:RESistance? configures the module to perform calibration of the internal calibration resistor and returns a 0 or 1 indicating the readiness of the module.

Parameters Returns: signed 16-bit integer (int16). 1 = ready, 0 = not ready.

Comments The internal calibration resistor should be calibrated before calibrating the channels in current mode.

This query configures the module for a 4-wire resistance measurement at the CAL output terminals when the calibration point is set to CALB (following a *RST or

DIAGnostic:CALibration:OUTPut[:MODE] CALBus command).

This query may also be used to set an external calibration resistor value when calibrating at each channel output.

This query is used together with the CALibration:VALue:RESistance command to calibrate the current calibration resistor.

Refer to Appendix D, “ Voltage/Current Output Adjustment” , starting on page 154, for additional information about this command.

Related Commands: CALibration:VALue:RESistance

CALibration:RESet

CALibration:RESet reloads, for module use, the calibration constants stored in non-volatile memory.

Parameters None.

Comments This command, with respect to the calibration constants, performs the same action as power-on.

At power-on, or following this command, calibration constants stored in non-volatile memory are loaded into the module where they are used during output operations.

If the trigger system is waiting for a trigger when this command is received, an error is generated.

This command opens all output relays.

Related Commands: CALibration:STORe

68 HP E1418A SCPI Command Reference

Chapter 3