SOURce

SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude]

SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <value MIN MAX DEF>

outputs the voltage set in value at channel n when the trigger conditions are satisfied and the channel output relay is closed. Use this command to synchronize multiple channel outputs.

Parameters

 

Parameter

Parameter

 

Range of

Default

 

Name

Type

 

Values

 

 

 

 

 

 

 

 

SOURcen

keyword

 

1 to 8 or 1 to 16

1

 

 

 

 

 

 

 

<value>

numeric (float 32) or

 

-16 V to +16 V or

0

 

 

discrete

 

MIN MAX DEF

 

 

 

 

 

 

Comments In the keyword, SOURcen, the valid range of n is:

 

 

1 to 8 for

E1418A, E1418A Option 002

 

 

 

1 to 16 for

E1418A Option 001, E1418A Option 003.

 

Values of n outside these ranges will generate an error.

The <value> parameter is a floating point 32-bit number in the range of -16 Volts to +16 Volts.

MIN sets -16 Volts. MAX sets +16 Volts. DEF sets 0 Volts.

This command sets channel n to output the voltage specified following a valid trigger condition. Following this command, the trigger system must be enabled using an INITiate command. See the TRIGger subsystem discussion, page 54, for additional information.

The channel output relay must be closed for the voltage to appear at the output terminals.

If the module is in a wait-for-trigger state (INIT has been received) when this command is received, an error is generated.

If the channel is in the current mode when this command is received, an error is generated.

Related Commands: SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?, SOURcen:FUNCtion:MODE VOLTage