SOURce
SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
SOURcen:VOLTage[:LEVel]:TRIGgered][:AMPLitude]? [<MIN MAX DEF>
queries the latest value written by either
SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude] or
SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude].
Parameters
Parameter | Parameter  | Range of  | Default  | 
Name  | Type  | Values  | 
  | 
  | 
  | 
  | 
  | 
SOURcen  | keyword  | 1 to 8 or 1 to 16  | 1  | 
  | 
  | 
  | 
  | 
<MIN MAX DEF>  | discrete (optional)  | MIN MAX DEF  | none  | 
  | 
  | 
  | 
  | 
Returns: 
Comments ∙ In the keyword, SOURcen, the valid range of n is:
1 to 8 for  | E1418A, E1418A Option 002  | 
1 to 16 for  | E1418A Option 001, E1418A Option 003.  | 
Values of n outside these ranges will generate an error.
∙If no parameter is given, this query returns the last value set on channel n by the
SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude] or
SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude] command. The value returned is the value last programmed on that channel, not an actual measurement of the hardware output value.
∙If a parameter is given,
MIN returns 
MAX returns +16.000.
DEF returns 0.000.
∙This command generates an error if channel n is in the current output mode.
∙Related Commands: SOURcen:VOLTage[:LEVel]:TRIGgered[:AMPLitude], SOURcen:FUNCtion:MODE VOLTage
94 HP E1418A SCPI Command Reference | Chapter 3  |