SOURce

SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude]

SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude] <value MIN MAX DEF>

immediately outputs the voltage set in value at channel n when the output relay is closed.

Parameters

 

Parameter

Parameter

 

Range of

Default

 

Name

Type

 

Values

 

 

 

 

 

 

 

 

SOURcen

keyword

 

1 to 8 or 1 to 16

1

 

 

 

 

 

 

 

<value>

numeric (float 32) or

 

-16 V to +16 V or

0

 

 

discrete

 

MIN MAX DEF

 

 

 

 

 

 

Comments In the keyword, SOURcen, the valid range of n is:

 

 

1 to 8 for

E1418A, E1418A Option 002

 

 

 

1 to 16 for

E1418A Option 001, E1418A Option 003.

 

Values of n outside these ranges will generate an error.

The <value> parameter is a floating point 32-bit number in the range of -16 Volts to +16 Volts.

MIN sets -16 Volts. MAX sets +16 Volts. DEF sets 0 Volts.

This command immediately programs the specified voltage on channel n. The output relay for that channel must be closed for the voltage to appear at the output terminals.

This command generates an error if channel n is in the current output mode.

Related Commands: SOURcen:FUNCtion:MODE VOLTage,

SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude]?, OUTPutn[:STATe]

Chapter 3

HP E1418A SCPI Command Reference 91