SOURce

SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude]

SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude] <value MIN MAX DEF>

immediately outputs the voltage set in value at channel n when the output relay is closed.

Parameters

 

Parameter

Parameter

 

Range of

Default

 

Name

Type

 

Values

 

 

 

 

 

 

 

 

SOURcen

keyword

 

1 to 8 or 1 to 16

1

 

 

 

 

 

 

 

<value>

numeric (float 32) or

 

-16 V to +16 V or

0

 

 

discrete

 

MIN MAX DEF

 

 

 

 

 

 

Comments ∙ In the keyword, SOURcen, the valid range of n is:

 

 

1 to 8 for

E1418A, E1418A Option 002

 

 

 

1 to 16 for

E1418A Option 001, E1418A Option 003.

 

Values of n outside these ranges will generate an error.

∙The <value> parameter is a floating point 32-bit number in the range of -16 Volts to +16 Volts.

MIN sets -16 Volts. MAX sets +16 Volts. DEF sets 0 Volts.

∙This command immediately programs the specified voltage on channel n. The output relay for that channel must be closed for the voltage to appear at the output terminals.

∙This command generates an error if channel n is in the current output mode.

∙Related Commands: SOURcen:FUNCtion:MODE VOLTage,

SOURcen:VOLTage[:LEVel][:IMMediate][:AMPLitude]?, OUTPutn[:STATe]

Chapter 3

HP E1418A SCPI Command Reference 91

Page 91
Image 91
HP E1418A manual SOURce SOURcenVOLTageLEVelIMMediateAMPLitude, SOURcenVOLTageLEVelIMMediateAMPLitude value MIN MAX DEF