![](/images/new-backgrounds/1330247/330247163x1.webp)
OUTPut
OUTPut:TTLTrgn[:STATe]?
OUTPut:TTLTrgn[:STATe]? queries if the specified VXIbus TTLTRG line is set to be a source of trigger from the module.
Parameters
Parameter | Parameter | Range of | Default |
Name | Type | Values |
|
|
|
|
|
TTLTrgn | keyword | 0 to 7 | none |
|
|
|
|
Returns: unsigned
Comments ∙ This query returns a 1 if the specified VXIbus TTLTRG line is enabled as a trigger source. This query returns a 0 if the specified VXIbus TTLTRG line is disabled as a trigger source.
∙Related Commands: OUTPut:TTLTn[:STATe]
OUTPutn[:STATe]
OUTPutn[:STATe] <state> opens or closes the specified channel output relay.
Parameters
| Parameter | Parameter |
| Range of | Default |
| Name | Type |
| Values |
|
|
|
|
|
|
|
| OUTPutn | keyword |
| 1 to 8 or 1 to 16 | none |
|
|
|
|
|
|
| <state> | boolean |
| OFF ON 0 1 | none |
|
|
|
|
| |
Comments ∙ In the keyword, OUTPutn, the valid range of n is: |
| ||||
| 1 to 8 for | E1418A, E1418A Option 002 |
|
| |
| 1 to 16 for | E1418A Option 001, E1418A Option 003. |
|
Values of n outside these ranges will generate an error.
∙This command sets the output state of individual channel output relays. Use the DIAGnostic:OUTPut:ALL[:STATe] command to set all channels with a single command.
∙*RST Condition: All output relays open.
∙Related Commands: OUTPutn[:STATe]?, DIAGnostic:OUTPut:ALL[:STATe]
82 HP E1418A SCPI Command Reference | Chapter 3 |