| Command Set | 5017/5018 |
| |
|
|
| Analog Input | |
| $aaSi1 |
|
|
|
| Name | Zero Calibration |
|
|
| Description Calibrates a specified analog input module to correct for | |||
|
| offset errors |
|
|
| Syntax | $aaSi1(cr) |
|
|
|
| $ is a delimiter character. |
|
|
|
| aa (range | ||
|
| address of the | ||
|
| calibrated. |
|
|
|
| Si identifies the I/O slot which is to be calibrated. | ||
|
| 1 represents the zero calibration command. | ||
|
| (cr) is the terminating character, carriage return (0Dh) | ||
| Response | !aa(cr) if the command is valid. | ||
|
| ?aa(cr) if an invalid operation was entered. | ||
|
| There is no response if the module detects a syntax error | ||
|
| or communication error or if the specified address does | ||
|
| not exist. |
|
|
|
| ! delimiter character indicating a valid command was | ||
|
| received. |
|
|
|
| ? delimiter character indicating the command was | ||
|
| invalid. |
|
|
|
| aa (range | ||
|
| address of an | ||
|
| (cr) is the terminating character, carriage return (0Dh) | ||
| Note: | In order to successfully calibrate an analog input | ||
|
| module's input range, a proper calibration input signal | ||
|
| should be connected to the analog input module before | ||
|
| and during the calibration process. (See also Chapter | ||
|
| 4, Section 4.5 on Calibration) | ||
|
|
|
|
|
|
| |||