% FS

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Advanced Features

An alternative method for measuring the output of the Current Loop converter is to connect a 250 ohm ±1% resistor across the current loop output in lieu of the current meter (see Figure 3-7for pin assignments and diagram of the analog output connector). This allows the use of a voltmeter connected across the resistor to measure converter output as VDC or mVDC.

V

+DC Gnd

Figure 7-7: Alternative Setup Using 250Ω Resistor for Checking Current Output Signal Levels

In this case, follow the procedure above but adjust the output for the following values:

Table 7-9: Current Loop Output Check

% FS

Voltage across

Voltage across

Resistor for 2-20 mA

Resistor for 4-20 mA

 

0

500 mVDC

1000 mVDC

100

5000 mVDC

5000 mVDC

04288D DCN5752

141

Page 143
Image 143
Teledyne M300EM operation manual Fs, Voltage across, Resistor for 2-20mA, Resistor for 4-20mA, mVDC