APPENDIX A-3: Warnings and Test Functions

APPENDIX A-3: Warnings and Test Functions

Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)

TEST FUNCTION NAME

MESSAGE TEXT

DESCRIPTION

COOFFSET2

OFFSET2=0.000

 

 

CO offset for range #2 in dual range

 

CO OFFSET2=0.000

7

mode, computed during zero/span

 

 

calibration.

 

 

 

 

CO2SLOPE 7

CO2 SLOPE=1.000

 

 

CO2 slope, computed during zero/span

 

 

 

 

calibration.

CO2OFFSET 7

CO2 OFFSET=0.000

 

CO2 offset, computed during zero/span

 

 

 

 

calibration.

O2SLOPE 10

O2 SLOPE=0.980

 

 

O2 slope, computed during zero/span

 

 

 

 

calibration.

O2OFFSET 10

O2 OFFSET=1.79 %

 

O2 offset, computed during zero/span

 

 

 

 

calibration.

CO

CO=17.7 PPM 3

 

 

CO concentration for current range.

CO2 7

CO2=15.0 %

 

 

CO2 concentration.

O2 10

O2=0.00 %

 

 

O2 concentration.

TESTCHAN

TEST=1751.4 MV

 

 

Value output to TEST_OUTPUT analog

 

 

 

 

output, selected with TEST_CHAN_ID

 

 

 

 

variable.

CLOCKTIME

TIME=09:52:20

 

 

Current instrument time of day clock.

RANGE(s)

User Configurable

 

 

CO2 range.

CO2RANGE

CO2 RANGE=20 % 1

 

O2RANGE

O2 RANGE=100 % 2

 

O2 range.

STABILITY

STABIL=0.0 PPM

 

 

Concentration stability (standard deviation

 

CO STB=0.0 PPM 1, 2

 

based on setting of STABIL_FREQ and

 

CO2 STB=0.0 % 1

 

 

STABIL_SAMPLES).

 

O2 STB=0.0 % 2

 

 

Detector measure reading.

COMEAS

CO MEAS=4125.0 MV

COREF

CO REF=3750.0 MV

 

Detector reference reading.

MRRATIO

MR RATIO=1.100

 

 

Measure/reference ratio.

SAMPPRESS

PRES=29.9 IN-HG-A

 

Sample pressure.

SAMPFLOW

SAMP FL=751 CC/M

 

Sample flow rate.

SAMPTEMP

SAMPLE TEMP=26.8 C

Sample temperature.

BENCHTEMP

BENCH TEMP=48.1 C

Bench temperature.

WHEELTEMP

WHEEL TEMP=68.1 C

Wheel temperature.

O2CELLTEMP 2

O2 CELL TEMP=50.2 C

O2 sensor cell temperature.

BOXTEMP

BOX TEMP=26.8 C

 

 

Internal chassis temperature.

PHOTOTEMP

PHT DRIVE=2500.0 MV

Photometer temperature.

COSLOPE

SLOPE=1.000

 

 

CO slope for current range, computed

 

CO SLOPE=1.000 1

 

 

during zero/span calibration.

COSLOPE1

SLOPE1=1.000

1

 

CO slope for range #1 in dual range

 

CO SLOPE1=1.000

 

mode, computed during zero/span

 

 

 

 

calibration.

COSLOPE2

SLOPE2=1.000

1

 

CO slope for range #2 in dual range

 

CO SLOPE2=1.000

 

mode, computed during zero/span

 

 

 

 

calibration.

COOFFSET

OFFSET=0.000

1

 

CO offset for current range, computed

 

CO OFFSET=0.000

 

during zero/span calibration.

COOFFSET1

OFFSET1=0.000

 

 

CO offset for range #1 in dual range

 

CO OFFSET1=0.000 1

mode, computed during zero/span

 

 

 

 

calibration.

COOFFSET2

OFFSET2=0.000

 

 

CO offset for range #2 in dual range

 

CO OFFSET2=0.000 1

mode, computed during zero/span

 

 

 

 

calibration.

CO2SLOPE 1

CO2 SLOPE=1.000

 

 

CO2 slope, computed during zero/span

 

 

 

 

calibration.

CO2OFFSET 1

CO2 OFFSET=0.000

 

CO2 offset, computed during zero/span

 

 

 

 

calibration.