Verify that the MODE

Getting Started

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

3.6.2.1.Verifying the M300E/EM Reporting Range Settings

While it is possible to perform the following procedure with any range setting we recommend that you perform this initial checkout using following reporting range settings:

Unit of Measure: PPM

Analog Output Reporting Range: 50 ppm

Mode Setting: SNGL

While these are the default setting for the M300E/EM Analyzer, it is recommended that you verify them before proceeding with the calibration procedure, by pressing:

SAMPLE

RANGE=50.0 PPM

CO= XX.XX

<TST

TST>

CAL

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

PRIMARY SETUP MENU

 

 

CFG

DAS RNGE PASS CLK MORE

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X RANGE CONTROL MENU

 

 

MODE SET

UNIT DIL

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Verify that the MODE

 

 

 

 

 

 

 

SETUP X.X

RANGE MODE:SINGL

 

 

is set for SNGL.

 

 

SNGL DUAL AUTO

 

 

 

ENTR EXIT

If it is not, press

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SINGL

ENTR.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X RANGE CONTROL MENU

 

 

 

 

 

 

MODE SET

UNIT

DIL

 

 

EXIT

Verify that the RANGE is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

set for 50.0

 

 

 

 

 

 

 

 

SETUP X.X

RANGE: 50.0 Conc

 

 

 

If it is not, toggle each

 

 

0

0

0

5

0

.0

ENTR EXIT

numeric key until the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

proper range is set, then

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

press ENTR.

 

 

 

 

 

 

 

 

 

 

SETUP X.X RANGE CONTROL MENU

 

 

 

 

 

 

 

 

 

 

 

 

MODE SET

UNIT

DIL

 

 

EXIT

Verify that the UNITs

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

CONC UNITS:PPM

 

 

 

is set for PPM

 

 

PPB

PPM

UGM MGM

 

 

ENTR EXIT

If it is not, press

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PPM

ENTR.

 

 

 

 

 

 

 

 

 

 

Press EXIT

3x’s to return the M200E to the SAMPLE mode.

52

04288D DCN5752

Page 54
Image 54
Teledyne M300EM is set for SNGL, If it is not, press, Singl, Entr, Verify that the RANGE is, press ENTR, is set for PPM