0) CONC: ATIMER, 1, 800

Advanced Features

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

7.1.5.SETUP DAS EDIT: ACCESSING THE IDAS EDIT MODE

iDAS configuration is most conveniently done through the APICOM remote control program. The following list of key strokes shows how to edit the iDAS using the front panel.

When editing the data channels, the top line of the display indicates some of the configuration parameters.

For example, the display line:

0) CONC: ATIMER, 1, 800

Translates to the following configuration:

Channel No.: 0

NAME: CONC

TRIGGER EVENT: ATIMER

PARAMETERS: One parameter is included in this channel

EVENT: This channel is set up to store 800 records.

To edit the name of a data channel, follow the above key sequence and refer to Section 7.1.5.1:

112

04288D DCN5752

Page 114
Image 114
Teledyne M300EM operation manual CONC: ATIMER, 1, Channel No.: NAME: CONC TRIGGER EVENT: ATIMER