Teledyne M300EM operation manual Internal Pneumatics OPT 50H

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 71
Image 71
5.6.4.1.Internal Pneumatics (OPT 50H)

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Optional Hardware and Software

5.6.4.ZERO/SPAN VALVE WITH INTERNAL CO SCRUBBER (OPTION 50H)

Option 50H is operationally and pneumatically similar to Option 50A above, except that the zero air is generated by an internal zero air scrubber. This means that the IZS inlet can simply be left open to ambient air.

Internal zero/span and sample/cal valves control the flow of gas through the instrument, but because the generator and calibrator limit the flow of zero air and span gas no shutoff valves are required.

5.6.4.1.Internal Pneumatics (OPT 50H)

Figure 5-7: Internal Pneumatic Flow OPT 50H – Zero/Span Valves with Internal Zero Air Scrubber

Table 5-3: Zero/Span Valve Operating States for Option 50H

MODE

VALVE

CONDITION

SAMPLE

Sample/Cal

Open to SAMPLE inlet

(Normal

 

 

Zero/Span

Open to ZERO AIR scrubber

State)

ZERO CAL

Sample/Cal

Open to ZERO/SPAN valve

Zero/Span

Open to ZERO AIR scrubber

 

SPAN CAL

Sample/Cal

Open to ZERO/SPAN valve

 

 

Zero/Span

Open to PRESSURE SPAN inlet

 

04288D DCN5752

69

Page 71
Image 71
Teledyne M300EM operation manual Internal Pneumatics OPT 50H