Teledyne M300EM Automatic Zero/Span Cal/Check Autocal, 2:AUTOCAL Modes, Mode Name, Action

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 200
Image 200

Calibration Procedures

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

9.3.3.3.Use of Zero/Span Valve with Remote Contact Closure

Contact closures for controlling calibration and calibration checks are located on the rear panel CONTROL IN connector. Instructions for setup and use of these contacts can be found in Section 3.3.4.

When the appropriate contacts are closed for at least 5 seconds, the instrument switches into zero, or span calibration mode and any internal zero/span valves installed will be automatically switched to the appropriate configuration.

The remote calibration contact closures may be activated in any order.

It is recommended that contact closures remain closed for at least 10 minutes to establish a reliable reading.

The instrument will stay in the selected mode for as long as the contacts remain closed.

If contact closures are being used in conjunction with the analyzer’s AutoCal (see Section 9.4) feature and the AutoCal attribute “CALIBRATE” is enabled, the M300E/EM will not recalibrate the analyzer until the contact is opened. At this point, the new calibration values will be recorded before the instrument returns to Sample Mode.

If the AutoCal attribute “CALIBRATE” is disabled, the instrument will return to Sample Mode, leaving the instrument’s internal calibration variables unchanged.

9.4.AUTOMATIC ZERO/SPAN CAL/CHECK (AUTOCAL)

The AutoCal system allows unattended periodic operation of the ZERO/SPAN valve options by using the M300E/EM Analyzer’s internal time of day clock. AutoCal operates by executing SEQUENCES programmed by the user to initiate the various calibration modes of the analyzer and open and close valves appropriately. It is possible to program and run up to three separate sequences (SEQ1, SEQ2 and SEQ3). Each sequence can operate in one of three modes, or be disabled.

 

Table 9-2: AUTOCAL Modes

 

 

MODE NAME

ACTION

DISABLED

Disables the Sequence.

ZERO

Causes the Sequence to perform a Zero calibration/check.

ZERO-SPAN

Causes the Sequence to perform a Zero point calibration/check followed by a Span

point calibration/check.

 

SPAN

Causes the Sequence to perform a Span concentration calibration/check only.

198

04288D DCN5752

Page 200
Image 200
Teledyne M300EM operation manual Automatic Zero/Span Cal/Check Autocal, 2:AUTOCAL Modes, Mode Name, Action