Table A-8:

Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)APPENDIX A-7: MODBUS Register Map

Table A-8:

Terminal Key Assignments

 

 

TERMINAL KEY ASSIGNMENTS

ESC

 

Abort line

CR (ENTER)

 

Execute command

Ctrl-C

 

Switch to computer mode

COMPUTER MODE KEY ASSIGNMENTS

LF (line feed)

 

Execute command

 

 

 

Ctrl-T

 

Switch to terminal mode

APPENDIX A-7: MODBUS Register Map

 

Table A-9: MODBUS Register Map

 

 

 

 

MODBUS

Description

Units

Register Address

 

 

(dec., 0-based)

 

 

 

MODBUS Floating Point Input Registers

 

(32-bit IEEE 754 format; read in high-word, low-word order; read-only)

0

Detector measure reading

mV

 

 

 

2

Detector reference reading

mV

 

 

 

4

M/R ratio.

none

 

 

 

6

Slope for range #1

none

8

Slope for range #2

none

10

Offset for range #1

none

12

Offset for range #2

none

14

Concentration for range #1 during zero/span calibration, just before

PPM

 

computing new slope and offset

 

16

Concentration for range #2 during zero/span calibration, just before

PPM

 

computing new slope and offset

 

18

Concentration for range #1

PPM

20

Concentration for range #2

PPM

 

 

 

22

Concentration stability

PPM

 

 

 

24

Bench temperature

°C

 

 

 

26

Bench temperature control duty cycle

Fraction

 

 

(0.0 = off,

 

 

1.0 = on full)

28

Wheel temperature

°C

30

Wheel temperature control duty cycle

Fraction

 

 

(0.0 = off,

 

 

1.0 = on full)

32

Sample temperature

°C

 

 

 

34

Sample pressure

“Hg

 

 

 

36

Internal box temperature

°C

 

 

 

38

Photometer detector temperature drive

mV

A-35

Page 357
Image 357
Teledyne M300E Modbus, Register Address, dec., 0-based, MODBUS Floating Point Input Registers, Description, Units, A-35