Teledyne M300EM Setup Variable, Numeric, Default, Value, Units, Range, Description, A-11

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 333
Image 333
A-11

Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)

APPENDIX A-2: Setup Variables For Serial I/O

Setup Variable

Numeric

Default

Value

 

Units

Value

Range

ET_TARGET_DET

mV

4375

0–5000

 

 

 

 

ET_TARGET_CONC

PPM

40,

1–9999.99

 

 

400 3, 8

 

ET_CONC_RANGE

Conc.

50,

0.1–50000

 

 

5000 3, 8

 

STD_TEMP

ºK

321

1–500

 

 

 

 

STD_PRESS

"Hg

28.5,

1–50

 

 

28.7 8,

 

 

 

28.8 12, 18,

 

 

 

28.1 4

 

BENCH_SET

ºC

48

0–100

 

 

Warnings:

 

 

 

43–53

 

 

 

 

 

WHEEL_SET

ºC

68,

0–100

 

 

62 19,23

 

 

 

Warnings:

 

 

 

63–73,

 

 

 

57–67 19,23

 

O2_CELL_SET 14

ºC

50

30–70

 

 

 

 

 

 

Warnings:

 

 

 

45–55

 

STD_O2_CELL_TEMP 14

ºK

323

1–500

 

 

 

 

ZERO_APPLY_IN_CAL 5

ON

OFF, ON

ZERO_DWELL 3, 5, 8

Seconds,

7,

1–60,

 

Minutes 5

3 5

1–30 5

 

 

 

 

ZERO_SAMPLES 3, 5, 8

Samples

15,

1–1000

 

 

750 5,

 

 

 

1000 19

 

ZERO_FILT_SIZE 3, 5, 8

Samples

5,

1–100

 

 

1 5

 

ZERO_LIMIT 3, 5, 8

Ratio

1.2,

0–5

 

 

1.15 3, 8,

 

 

 

1 5

 

ZERO_CAL 3, 5, 8

Ratio

1.18

0.5–5

Description

Target detector reading during electrical test.

Target concentration during electrical test.

D/A concentration range during electrical test.

Standard temperature for temperature compensation.

Standard pressure for pressure compensation.

Optical bench temperature set point and warning limits.

Wheel temperature set point and warning limits.

O2 sensor cell temperature set point and warning limits.

Standard O2 cell temperature for temperature compensation.

ON applies auto-reference offset and dilution factor during zero/span calibration;

OFF disables both.

(Only applicable if ZERO_ENABLE is ON.)

Dwell time after closing or opening zero scrubber valve.

Number of zero samples to average.

Auto-zero offset moving average filter size.

Minimum auto-zero ratio allowed; must be greater than this value to be valid.

Calibrated auto-zero ratio.

A-11

Page 333
Image 333
Teledyne M300EM operation manual Setup Variable, Numeric, Default, Value, Units, Range, Description, A-11