Table A-2: T300/T300M and M300E/EM Warning Messages, Revision L.8

 

 

Name 1

Message Text

 

Warnings

WSYSRES

SYSTEM RESET

 

 

WDATAINIT

DATA INITIALIZED

WCONFIGINIT

CONFIG INITIALIZED

 

 

WCONCALARM1

CONC ALARM 1 WARN

WCONCALARM2

CONC ALARM 2 WARN

WSOURCE

SOURCE WARNING

 

 

WAUTOZERO 4, 5

AZERO WARN 1.001

 

 

WBENCHTEMP

BENCH TEMP WARNING

 

 

WWHEELTEMP

WHEEL TEMP WARNING

 

 

WO2CELLTEMP 10

O2 CELL TEMP WARN

 

 

WSAMPFLOW 6

SAMPLE FLOW WARN

 

 

WSAMPPRESS

SAMPLE PRESS WARN

 

 

WSAMPTEMP

SAMPLE TEMP WARN

 

 

WPURGEPRESS 9

PURGE PRESS WARN

 

 

WBOXTEMP

BOX TEMP WARNING

 

 

WBOXTEMP2 4

BOX TEMP2 WARNING

 

 

WOVENTEMP 11

OVEN TEMP WARNING

 

 

WPHOTOTEMP

PHOTO TEMP WARNING

 

 

WDYNZERO

CANNOT DYN ZERO

 

 

WDYNSPAN

CANNOT DYN SPAN

 

 

WREARBOARD

REAR BOARD NOT DET

 

 

Description

Instrument was power-cycled or the CPU was reset.

Data storage was erased.

Configuration storage was reset to factory configuration or erased.

Concentration limit 1 exceeded. Concentration limit 2 exceeded.

Reference reading minus dark offset outside of warning limits specified by

REF_SOURCE_LIMIT variable.

Auto-reference ratio below limit specified by ZERO_LIMIT variable.

Bench temperature outside of warning limits specified by BENCH_SET variable.

Wheel temperature outside of warning limits specified by WHEEL_SET variable.

O2 sensor cell temperature outside of warning limits specified by O2_CELL_SET variable.

Sample flow outside of warning limits specified by SAMP_FLOW_SET variable.

Sample pressure outside of warning limits specified by SAMP_PRESS_SET variable.

Sample temperature outside of warning limits specified by SAMP_TEMP_SET variable.

Purge pressure outside of warning limits specified by PURGE_PRESS_SET variable.

Internal box temperature outside of warning limits specified by BOX_SET variable.

Internal box temperature #2 outside of warning limits specified by BOX2_SET variable.

Oven temperature outside of warning limits specified by OVEN_SET variable.

Photometer temperature outside of warning limits specified by PHOTO_TEMP_SET variable.

Contact closure zero calibration failed while DYN_ZERO was set to ON.

Contact closure span calibration failed while DYN_SPAN was set to ON.

Rear board was not detected during power up.