Teledyne M300EM VR1 Adjustment Made Here, Housing Mounting Screws Sync/Demod PCA Housing

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 305
Image 305
Housing Mounting

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Troubleshooting & Repair

13.6.3.2.Adjusting the Sync/Demodulator, Circuit Gain

To adjust the sync/demodulator circuit gain:

1.Make sure that the analyzer is turned on and warmed up.

2.Set the analyzer display to show the STABIL or CO STB test function.

3.Apply Zero Air to Sample Inlet of the analyzer.

4.Wait until the stability reading falls below 1.0 ppm.

5.Change the analyzer display to show the CO MEAS.

6.Remove the Sync/Demod Housing

Remove the two mounting screws.

Carefully lift the housing to reveal the sync/demod PCA.

Housing Mounting

Screws

 Optical Bench Sync/Demod

PCA Housing

VR1 Optical Bench

Figure 13-19: Location of Sync/Demod Housing Mounting Screws

7. Adjust potentiometer VR1 until CO MEAS reads 4500 mV ± 300 mV

VR1

Adjustment Made Here

Figure 13-20: Location of Sync/Demod Gain Potentiometer

04288D DCN5752

303

Page 305
Image 305
Teledyne M300EM operation manual VR1 Adjustment Made Here, Housing Mounting Screws Sync/Demod PCA Housing, Optical Bench