Teledyne M300EM operation manual ∙Leaks ∙Correct flow

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 227
Image 227
∙Leaks

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

EPA Calibration Protocol

SAMPLE

A1:CONC1=50 PPM

CO = XXXX

< TST TST > CAL

SETUP

 

 

 

 

 

 

SAMPLE

CO STB=XXXX PPB

CO=XXXX

< TST TST > CAL

SETUP

 

 

 

ACTION:

Allow calibration gas diluted to proper concentration for

Midpoint N to enter the sample port

Set the Display to show the

COSTB test function.

This function calculates the

stability of the CO

measurement.

Wait until

STABIL falls

below 0.2 PPM

(for M300E). This may take several minutes.

SAMPLE

COSTB=XXXX PPB

CO=XXXX

 

< TST TST > CAL CALZ CALS

SETUP

 

 

 

 

Record the CO

 

 

 

 

 

 

reading as

SPAN CAL M

A1:CONC1=50 PPM

CO = XXXX

displayed on the

< TST TST > ZERO SPAN CONC

EXIT

instrument’s front

panel.

 

 

 

 

Press EXIT to

Return to the

Main SAMPLE

Display.

ACTION:

Allow Calibration Gas diluted to

proper concentration for

Midpoint N+1 to enter the sample

port.

Plot the analyzer responses versus the corresponding calculated concentrations to obtain a calibration relationship. Determine the best-fit straight line (y = mx + b) determined by the method of least squares.

After the best-fit line has been drawn, determine whether the analyzer response is linear. To be considered linear, no calibration point should differ from the best-fit line by more than ±2% of full scale.

If carried out carefully, the checks described in this section will provide reasonable confidence that the M300E is operating properly. Checks should be carried out at least every 3 months as the possibility of malfunction is always present.

If the linearity error is excessive and cannot be attributed to outside causes, check the M300E system for:

Sample pressure higher than ambient – pressurized sample gas

Leaks

Correct flow

Miscalibrated span gas tanks or bad zero gas

Miscalibrated sample pressure transducer

Failed IR detector, GFC Wheel or Sync/Demod Board

Contaminated optical bench or sample lines