Teledyne M300EM Before Calibration, Calibration Gases, Zero Air, troubleshooting tips

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 188
Image 188
9.1.BEFORE CALIBRATION

Calibration Procedures

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

9.1.BEFORE CALIBRATION

The calibration procedures in this section assume that the range mode, analog range and units of measure have already been selected for the analyzer. If this has not been done, please do so before continuing (see Section 6.6 for instructions).

NOTE

If any problems occur while performing the following calibration procedures, refer to Section 12 for

troubleshooting tips.

9.1.1.REQUIRED EQUIPMENT, SUPPLIES, AND EXPENDABLES

Calibration of the M300E/EM Analyzer requires a certain amount of equipment and supplies. These include, but are not limited to, the following:

Zero-air source.

Span gas source.

Gas lines - All Gas lines should be PTFE (Teflon), FEP, glass, stainless steel or brass.

A recording device such as a strip-chart recorder and/or data logger (optional). For electronic documentation, the internal data acquisition system iDAS can be used.

NOTE

If any problems occur while performing the following calibration procedures, refer to Section 12 of this

manual for troubleshooting tips.

9.1.2.CALIBRATION GASES

9.1.2.1.Zero Air

Zero air or zero calibration gas is defined as a gas that is similar in chemical composition to the measured medium but without the gas to be measured by the analyzer.

For the M300E/EM zero air should contain less than 25 ppb of CO and other major interfering gases such as CO and Water Vapor. It should have a dew point of -5°C or less.

If your application is not a measurement in ambient air, the zero calibration gas should be matched to the composition of the gas being measured.

Pure nitrogen (N2) can be used as a zero gas for applications where CO is measured in nitrogen.

If your analyzer is equipped with an external zero air scrubber option, it is capable of creating zero air from ambient air.

For analyzers without the zero air scrubber, a zero air generator such as the Teledyne API’s M701 can be used. Please visit the company website for more information.

186

04288D DCN5752

Page 188
Image 188
Teledyne M300EM operation manual Before Calibration, Calibration Gases, Zero Air, manual for troubleshooting tips