Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)

APPENDIX A-2: Setup Variables For Serial I/O

Setup Variable

Numeric

Default

Value

 

Units

Value

Range

TEST_CHAN_ID

NONE 0

NONE,

 

 

 

CO

 

 

 

MEASURE,

 

 

 

CO

 

 

 

REFERENC

 

 

 

E,

 

 

 

VACUUM

 

 

 

PRESSURE,

 

 

 

SAMPLE

 

 

 

PRESSURE,

 

 

 

SAMPLE

 

 

 

FLOW,

 

 

 

SAMPLE

 

 

 

TEMP,

 

 

 

BENCH

 

 

 

TEMP,

 

 

 

WHEEL

 

 

 

TEMP,

 

 

 

O2 CELL

 

 

 

TEMP 14,

 

 

 

CHASSIS

 

 

 

TEMP,

 

 

 

PHT DRIVE,

 

 

 

TEMP4 5

REMOTE_CAL_MODE

LOW 0

LOW,

 

 

 

HIGH,

 

 

 

CO2 10,

 

 

 

O2 14

PASS_ENABLE

OFF

ON, OFF

 

 

 

 

STABIL_FREQ

Seconds

10

1–300

 

 

120 19, 23

 

STABIL_SAMPLES

Samples

25

2–40

 

 

 

 

PHOTO_TEMP_SET

mV

2500

0–5000

 

 

Warnings:

 

 

 

250–4750

 

 

 

 

 

SAMP_PRESS_SET

In-Hg

29.92

0–100

 

 

Warnings:

 

 

 

15–32

 

Description

Diagnostic analog output ID.

CO range or other gas to calibrate during contact closure or Hessen calibration.

ON enables passwords; OFF disables them.

Stability measurement sampling frequency.

Number of samples in concentration stability reading.

Photometer temperature warning limits. Set point is not used.

Sample pressure warning limits. Set point is not used.