6.BASIC OPERATION

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Basic Operation

6.BASIC OPERATION

The M300E/EM Analyzer is a computer-controlled analyzer with a dynamic menu interface that allows all major operations to be controlled from the front panel display and keyboard through user-friendly menus (a complete set of menu trees is located in Appendix A of this manual).

This section includes step-by-step instructions for using the display/keyboard to set up and operate the M300E/EM Analyzer's basic CO measurement features and functional modes.

6.1.OVERVIEW OF OPERATING MODES

The M300E/EM software has a variety of operating modes. Most commonly, the analyzer will be operating in Sample Mode. In this mode a continuous read-out of the CO concentration is displayed on the front panel. If the analyzer is configured to measure more than one gas (e.g. CO along with O2 or CO2) the display will cycle through gas list.

While in SAMPLE mode calibrations can be performed and TEST functions as well as WARNING messages can be examined. If any of the analyzer’s analog outputs are enabled, the current concentration value will be available at the analog output connector.

The second most important operating mode is SETUP mode. This mode is used for performing certain configuration operations, such as programming the iDAS system or the configurable analog output channels, or setting up the analyzer’s serial communication channels (RS-232/RS-485/Ethernet). The SETUP mode is also used for performing various diagnostic tests during troubleshooting.

Message Field

Mode Field

Concentration Field

SAMPLE

 

RANGE=50.00 PPM

 

CO= XX.XX

<TST TST> CAL

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 6-1: Front Panel Display

The mode field of the front panel display indicates to the user which operating mode the unit is currently running. Besides SAMPLE and SETUP, other modes the analyzer can be operated in are:

Table 6-1: Analyzer Operating Modes

MODE

SAMPLE

SAMPLE A

M-P CAL

SETUP [X.X]

ZERO CAL [type] 1,2 & 3

LO CAL [type] 2 & 3

SPAN CAL [type] 1,2 & 3

DIAG Mode

EXPLANATION

Sampling normally, flashing text indicates adaptive filter is on.

Indicates that unit is in Sample Mode while AUTOCAL feature is active (Internal Span Only). This is the basic calibration mode of the instrument and is activated by pressing the CAL key.

SETUP mode is being used to configure the analyzer. The gas measurement will continue during this process. The revision of the M300E/EM firmware being run will appear after the word “SETUP

Unit is performing ZERO calibration procedure.

Unit is performing LOW SPAN (midpoint) cal check procedure. Unit is performing SPAN calibration procedure.

One of the analyzer’s diagnostic modes is active (Section 7.3).

[type:]

1A: Initiated automatically by the AUTOCAL feature (Internal Span Only).

2M: initiated manually by the user via the front panel controls.

3R: initiated remotely through the COM ports or digital control inputs.

04288D DCN5752

85

Page 87
Image 87
Teledyne M300EM operation manual Basic Operation, Overview Of Operating Modes, Explanation, 1:Front Panel Display