Teledyne M300EM Setup Variable, Numeric, Default, Value, Units, Range, Description, A-13

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 335
Image 335
A-13

Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)

APPENDIX A-2: Setup Variables For Serial I/O

Setup Variable

Numeric

Default

Value

 

Units

Value

Range

RANGE_MODE

SNGL 0

SNGL,

 

 

 

DUAL,

 

 

 

AUTO

 

 

 

 

CONC_RANGE1

Conc.

50,

0.1–50000

 

 

200 6,

 

 

 

500 3, 8

 

CONC_RANGE2 1

Conc.

50,

0.1–50000

 

 

200 6,

 

 

 

500 3, 8

 

CO2_RANGE 10

%

15

0.1–500,

 

 

 

0.1–2000 16

O2_RANGE 14

%

100

0.1–500

RS232_MODE

BitFlag

0

0–65535

Description

Range control mode.

D/A concentration range 1.

D/A concentration range 2.

CO2 concentration range.

O2 concentration range.

RS-232 COM1 mode flags. Add values to combine flags.

1 = quiet mode

2 = computer mode

4 = enable security

8 = enable hardware handshaking

16 = enable Hessen protocol 11

32 = enable multi-drop

64 = enable modem

128 = ignore RS-232 line errors

256 = disable XON / XOFF support

512 = disable hardware FIFOs

1024 = enable RS-485 mode

2048 = even parity, 7 data bits, 1 stop bit

4096 = enable command prompt

8192 = even parity, 8 data bits, 1 stop bit

16384 = enable dedicated MODBUS ASCII protocol

32678 = enable dedicated MODBUS RTU or TCP protocol

BAUD_RATE

115200 0

300,

1200,

2400,

4800,

9600,

19200,

38400,

57600,

115200

16384 = enable TAI protocol 17 RS-232 COM1 baud rate.

A-13

Page 335
Image 335
Teledyne M300EM operation manual Setup Variable, Numeric, Default, Value, Units, Range, Description, A-13