7.5.SETUP MORE ALRM: USING THE GAS CONCENTRATION ALARMS

The M300E/EM includes two CO concentration alarms if OPT 61 is installed on your instrument. Each alarm has a user settable limit, and is associated with a Single Pole Double Throw relay output accessible via the alarm output connector on the instrument’s back panel (See Section 3.3.3). If the CO concentration measured by the instrument rises above that limit, the alarm‘s status output relay is closed.

The default settings for ALM1 and ALM2 are:

Table 7-11: CO Concentration Alarm Default Settings

ALARM

STATUS

LIMIT SET POINT1

alm1

Disabled

100 ppm

alm2

Disabled

300 ppm

1Set points listed are for PPM. Should the reporting range units of measure be changed (See Section 6.6.3) the analyzer will automatically scale the set points to match the new range unit setting.

NOTE

To prevent the concentration alarms from activating during span calibration operations ensure that the

CAL or CALS button is pressed prior to introducing span gas into the analyzer.